assemblies
Source, normal, reverse, HI-open, LO-open, short and reference-DMM pairs.
HR-30 / off-robot metrology / FER-G11 preparation
One blue normal cable. Four physically segregated fault tools. A manufacturer-built source pair and DMM patch pair. Every conductor, empty contact, strip length, ferrule and torque step is controlled.
Source, normal, reverse, HI-open, LO-open, short and reference-DMM pairs.
Each contact is mapped; intentional opens are physically empty.
Phoenix terminal torque; actual values and tool identity must be recorded.
Parts, crimp coupons, insulation limit and qualified review remain open.
CK-02 is blue and straight-through. It is the only transfer cable used for ordinary point calibration.
CK-03 through CK-06 are red, orange or black/yellow, kept in a separate compartment and removed after each controlled case.
No plug, adapter or meter lead is fitted, removed or rerouted with the source output enabled.
| Assembly Id | Name | Function | Construction | Finished Length Mm | End A | End B | Label Color | Maximum Use | State | Warning |
|---|---|---|---|---|---|---|---|---|---|---|
| CK-01 | isolated source lead pair | Keysight output 2 or 3 to fixture JPS | two manufacturer-assembled Mueller 18 AWG silicone leads; factory tin-dipped ends removed and re-terminated | 975 target; record actual; 950 minimum | BU-0061-M-39-2 red and -0 black shrouded 4 mm bananas | one Phoenix 1757019, contact 1 red HI and contact 2 black LO | WHITE body / RED HI / BLACK LO | 24 V DC candidate, 10 mA source-current-limit candidate | UNBUILT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-02 | normal transfer cable | fixture JDUT to selected pod JIN | Alpha 5610B2201, both conductors populated straight through; shield and drain isolated at both ends | 300 +/-5 jacket span | Phoenix 1757019 JDUT | Phoenix 1757019 POD JIN | BLUE - NORMAL ONLY | 24 V DC candidate, one disconnected chain | UNBUILT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-03 | reversed-polarity adapter | controlled negative-response characterization | Alpha 5610B2201, contacts crossed; shield and drain isolated at both ends | 300 +/-5 jacket span | Phoenix 1757019 JDUT | Phoenix 1757019 POD JIN | RED - REVERSE POLARITY | 24 V DC candidate only under CF-P10 | UNBUILT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-04 | HI-open adapter | controlled CAL_HI open characterization | Alpha 5610B2201; black LO conductor only; white conductor and shield/drain cut back and insulated | 300 +/-5 jacket span | Phoenix 1757019 JDUT; contact 1 physically empty | Phoenix 1757019 POD JIN; contact 1 physically empty | ORANGE - HI OPEN | 24 V DC candidate only under CF-P11 | UNBUILT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-05 | LO-open adapter | controlled CAL_LO open characterization | Alpha 5610B2201; white HI conductor only; black conductor and shield/drain cut back and insulated | 300 +/-5 jacket span | Phoenix 1757019 JDUT; contact 2 physically empty | Phoenix 1757019 POD JIN; contact 2 physically empty | ORANGE - LO OPEN | 24 V DC candidate only under CF-P11 | UNBUILT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-06 | current-limit shorting adapter | controlled CAL_HI-to-CAL_LO short at JDUT | one Phoenix 1757019 with an 80 +/-5 mm 22 AWG insulated loop; two ferrules | 80 +/-5 conductor before termination | Phoenix 1757019 contact 1 to contact 2 | NONE - capped single-ended adapter | BLACK/YELLOW - SHORT; OUTPUT OFF TO FIT/REMOVE | candidate 10 mA current-limit characterization only | UNBUILT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-07 | reference-DMM patch pair | fixture JHI/JLO to Fluke V/ohm and COM | unaltered manufacturer-assembled Fluke TL930 red/black pair | 610 manufacturer nominal | red/black multi-stacking 4 mm plugs to JHI/JLO | red to V/ohm, black to COM; both current jacks empty | GREEN - REFERENCE DMM ONLY | 30 V RMS / 60 V DC, 8 A manufacturer rating; fixture candidate <=24 V DC | UNBUILT / RECEIVE AND INSPECT | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| Assembly Id | Circuit | From Connector | From Contact | To Connector | To Contact | Signal Or Fault | Conductor | Shield Or Drain | Continuity Target | State | Warning |
|---|---|---|---|---|---|---|---|---|---|---|---|
| CK-01 | 1 | KEYSIGHT + RED | 4 mm plug | JPS | 1 | CAL_HI | red 18 AWG | NOT PRESENT | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-01 | 2 | KEYSIGHT - BLACK | 4 mm plug | JPS | 2 | CAL_LO | black 18 AWG | NOT PRESENT | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-02 | 1 | JDUT | 1 | POD JIN | 1 | CAL_HI | white 22 AWG | CUT BACK AND INSULATED BOTH ENDS | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-02 | 2 | JDUT | 2 | POD JIN | 2 | CAL_LO | black 22 AWG | CUT BACK AND INSULATED BOTH ENDS | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-03 | 1 | JDUT | 1 | POD JIN | 2 | CROSSED HI TO LO | white 22 AWG | CUT BACK AND INSULATED BOTH ENDS | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-03 | 2 | JDUT | 2 | POD JIN | 1 | CROSSED LO TO HI | black 22 AWG | CUT BACK AND INSULATED BOTH ENDS | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-04 | 1 | JDUT | 1 EMPTY | POD JIN | 1 EMPTY | CONTROLLED HI OPEN | NO CONDUCTOR | CUT BACK AND INSULATED BOTH ENDS | >10 Mohm; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-04 | 2 | JDUT | 2 | POD JIN | 2 | CAL_LO | black 22 AWG | CUT BACK AND INSULATED BOTH ENDS | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-05 | 1 | JDUT | 1 | POD JIN | 1 | CAL_HI | white 22 AWG | CUT BACK AND INSULATED BOTH ENDS | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-05 | 2 | JDUT | 2 EMPTY | POD JIN | 2 EMPTY | CONTROLLED LO OPEN | NO CONDUCTOR | CUT BACK AND INSULATED BOTH ENDS | >10 Mohm; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-06 | 1 | JDUT | 1 | JDUT | 2 | CONTROLLED HI-LO SHORT | 22 AWG offcut loop | NOT PRESENT | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-07 | 1 | JHI RED | 4 mm jack | FLUKE | V/OHM | CAL_HI REFERENCE | red TL930 | NOT PRESENT | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-07 | 2 | JLO BLACK | 4 mm jack | FLUKE | COM | CAL_LO REFERENCE | black TL930 | NOT PRESENT | <1 ohm after lead-null; actual recorded | NOT EXECUTED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| Cut Id | Assembly Id | Material | Quantity | Raw Length Mm | Finished Length Mm | End Preparation | Tolerance Or Gate | Warning |
|---|---|---|---|---|---|---|---|---|
| CUT-01 | CK-01 | Mueller BU-0061-M-39-2 red | 1 | 990.6 nominal manufacturer assembly | 975 target; 950 minimum | retain banana end; remove complete tin-dipped segment; strip 9 mm; crimp 3200742 | record actual; reject below 950 mm or with insulation damage | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-02 | CK-01 | Mueller BU-0061-M-39-0 black | 1 | 990.6 nominal manufacturer assembly | 975 target; 950 minimum | retain banana end; remove complete tin-dipped segment; strip 9 mm; crimp 3200742 | record actual; reject below 950 mm or with insulation damage | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-03 | CK-02 | Alpha Wire 5610B2201 | 1 | 360 | 300 +/-5 jacket span | remove 30 +/-1 mm jacket each end; cut foil/drain flush; sleeve 20 +/-2 mm FIT-221-1/4; strip populated conductors 10 mm; crimp 3203066 | no shield strand, nicked conductor or exposed copper; empty contacts physically unpopulated | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-04 | CK-03 | Alpha Wire 5610B2201 | 1 | 360 | 300 +/-5 jacket span | remove 30 +/-1 mm jacket each end; cut foil/drain flush; sleeve 20 +/-2 mm FIT-221-1/4; strip populated conductors 10 mm; crimp 3203066 | no shield strand, nicked conductor or exposed copper; empty contacts physically unpopulated | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-05 | CK-04 | Alpha Wire 5610B2201 | 1 | 360 | 300 +/-5 jacket span | remove 30 +/-1 mm jacket each end; cut foil/drain flush; sleeve 20 +/-2 mm FIT-221-1/4; strip populated conductors 10 mm; crimp 3203066 | no shield strand, nicked conductor or exposed copper; empty contacts physically unpopulated | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-06 | CK-05 | Alpha Wire 5610B2201 | 1 | 360 | 300 +/-5 jacket span | remove 30 +/-1 mm jacket each end; cut foil/drain flush; sleeve 20 +/-2 mm FIT-221-1/4; strip populated conductors 10 mm; crimp 3203066 | no shield strand, nicked conductor or exposed copper; empty contacts physically unpopulated | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-07 | CK-06 | one insulated 22 AWG conductor from 5610B2201 controlled offcut | 1 | 100 | 80 +/-5 before termination | strip 10 mm each end; crimp one 3203066 per end; form loop without sharp bend | minimum inside bend radius 10 mm; no exposed copper | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CUT-08 | CK-07 | Fluke TL930 part 1616671 | 1 | 610 nominal | DO NOT CUT OR MODIFY | receive, inspect, label only | reject damaged insulation, loose plug, wrong colors or failed continuity | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| Step Id | Operation | Instruction | Required Record | Stop Rule | Result | Authority | Warning |
|---|---|---|---|---|---|---|---|
| TR-01 | kit segregation | Clear bench; no robot, battery, safety circuit, actuator, USB-grounded equipment or live source connected. | Photograph empty bench and fixture isolation | STOP on any robot or energized circuit connection | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-02 | incoming inspection | Record manufacturer, order code, lot and quantity for every cable, connector, ferrule, shrink and tool. | Incoming record and photographs | STOP on substitution or damage | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-03 | cut and strip | Process one assembly at a time from wire-cut-list.csv; use calibrated length and stripping gauges. | Actual lengths and strip measurements | STOP on nicked strands or tolerance failure | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-04 | shield isolation | For CK-02..05 cut foil/drain flush at both ends and cover with 20 +/-2 mm FIT-221-1/4 sleeve. | No shield/drain continuity to any contact | STOP on exposed shield strand | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-05 | ferrule crimp | Use CRIMPFOX CENTRUS 10S: 3200742 on 18 AWG source leads; 3203066 on 22 AWG conductors. | Lot/tool ID, visual crimp inspection and coupon pull record | STOP on wrong ferrule/tool or incomplete insertion | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-06 | plug termination | Insert to contact map; tighten every Phoenix 1757019 screw to 0.50-0.60 N m with in-calibration torque driver. | Contact, torque value, driver serial/date | STOP on loose strand, exposed copper or torque miss | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-07 | labels | Apply both-end assembly ID, role, direction and fault label; orange open adapters name the open conductor; short adapter states OUTPUT OFF TO FIT/REMOVE. | Two-person label-to-map check | STOP on ambiguity or unreadable text | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-08 | continuity and isolation | Four-wire/lead-null continuity per populated path; >10 Mohm between open contacts, shield/drain and populated circuits at a qualified test voltage not yet released. | Raw readings and instrument identities | STOP until insulation test voltage is selected | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-09 | polarity fixture | With no source connected, use continuity to verify CK-02 straight, CK-03 crossed, CK-04 HI open, CK-05 LO open and CK-06 short. | Signed connector/contact matrix | STOP on any mismatch | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-10 | retention and bend | Perform qualified axial retention coupon/test and verify minimum 49 mm cable bend radius for 5610B2201 assemblies. | Force, duration, displacement and post-test continuity | STOP because numeric retention limit remains open | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-11 | DMM pair | Verify CK-07 red JHI-to-V/ohm and black JLO-to-COM; meter current jacks remain visibly empty. | Photograph and continuity readings | STOP on current-jack use | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| TR-12 | bag and control | Cap contacts; bag each assembly separately; store CK-03..06 in FAULT ADAPTER compartment physically separated from CK-02. | Inventory, seal and custodian | No adapter may be left attached | NOT EXECUTED | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| Test Id | Test | Acceptance Or Evidence | Result | Review | Warning |
|---|---|---|---|---|---|
| CK-T01 | identity and count | all seven assemblies present; order codes and lots recorded | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T02 | contact population | every populated and intentionally empty contact matches connector-contact-map.csv | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T03 | continuity | every populated path <1 ohm after lead-null; actual readings retained; final numeric limit qualified | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T04 | open isolation | intended opens and shield/drain >10 Mohm at qualified test voltage; voltage selection remains open | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T05 | polarity | CK-01/02/07 straight; CK-03 crossed; CK-04/05 single-conductor; CK-06 short only | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T06 | screw torque | every 1757019 terminal 0.50-0.60 N m with in-calibration driver; actual value recorded | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T07 | crimp coupons | 18 AWG/3200742 and 22 AWG/3203066 visual and pull coupons accepted to a released numeric limit | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T08 | strain and bend | no conductor load at terminal; 5610B2201 bend radius >=49 mm in use | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T09 | labels | fault adapters recognizable from either end and segregated from blue normal cable | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-T10 | low-energy dry rehearsal | source absent; correct cable selected from procedure and returned after each simulated step | NOT EXECUTED | REQUIRED | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| Item Id | Quantity | Unit | Manufacturer | Order Code | Description | Selection State | Procurement Released | Warning |
|---|---|---|---|---|---|---|---|---|
| CK-B01 | 1 | each | Mueller Electric | BU-0061-M-39-2 | red 39 inch 18 AWG silicone shrouded-banana lead | EXACT CANDIDATE | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B02 | 1 | each | Mueller Electric | BU-0061-M-39-0 | black 39 inch 18 AWG silicone shrouded-banana lead | EXACT CANDIDATE | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B03 | 2 | m | Alpha Wire | 5610B2201 | one-pair shielded 22 AWG cable | EXACT CANDIDATE; includes process scrap | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B04 | 10 | each | Phoenix Contact | 1757019 | two-position 5.08 mm screw plug | EXACT CANDIDATE | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B05 | 4 | each | Phoenix Contact | 3200742 | AI 1-6 RD ferrule for 18 AWG source leads plus two coupons/spares | EXACT CANDIDATE | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B06 | 20 | each | Phoenix Contact | 3203066 | AI 0.34-8 TQ ferrule for 22 AWG assemblies plus coupons/spares | EXACT CANDIDATE | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B07 | 1 | each | Phoenix Contact | 1213154 | CRIMPFOX CENTRUS 10S hand crimper | EXACT TOOL CANDIDATE; calibration/inspection open | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B08 | 1 | each | Alpha Wire | FIT-KIT-221BK | black FIT-221 heat-shrink kit containing 1/4 inch tubing | EXACT MATERIAL CANDIDATE; shrink process open | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B09 | 1 | each | Fluke | TL930 / part 1616671 | 61 cm red/black 4 mm patch-cord pair | EXACT CANDIDATE; do not modify | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B10 | 1 | each | SELECTION REQUIRED | SELECTION REQUIRED | calibrated 0.50-0.60 N m torque driver and bit | SELECTION REQUIRED | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B11 | 1 | each | PROJECT PRINT | CK-LABEL-P0.1 | water-resistant two-ended cable and fault labels | MATERIAL/ADHESION SELECTION REQUIRED | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-B12 | 1 | each | SELECTION REQUIRED | SELECTION REQUIRED | latching divided storage case with normal/fault physical segregation | SELECTION REQUIRED | NO | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| Hold Id | Open Item | Closure Evidence | State | Authority | Warning |
|---|---|---|---|---|---|
| CK-H01 | received parts and lots | supplier records, identity and incoming inspection | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H02 | source-lead tin removal and final length | released work instruction plus actual lengths >=950 mm | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H03 | crimp pull limits | qualified 18 AWG and 22 AWG coupon values, sample count and results | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H04 | insulation test voltage and limit | qualified low-energy test method compatible with connected components | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H05 | torque tool | exact calibrated driver/bit selection and in-date certificate | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H06 | label and storage materials | durability/adhesion and physical fault-adapter segregation | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H07 | fixture and received-mating FAI | 73099/TL930 fit, 1757019 mating, enclosure access and bend clearance | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H08 | fabrication and inspection | signed traveler and all CK-T01..10 evidence | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H09 | qualified electrical/metrology review | accept contact maps, tests and calibration-procedure use | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
| CK-H10 | FER-G11 and work authority | separate signed staged authorization; this kit cannot grant it | OPEN | NO PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED-ROBOT-TEST, MOTION, WALKING OR ENERGIZATION AUTHORITY | PRELIMINARY - UNBUILT OFF-ROBOT CALIBRATION CABLE AND FAULT-ADAPTER KIT - ZERO SAFETY CREDIT - NOT APPROVED FOR PROCUREMENT, FABRICATION, ASSEMBLY, ROBOT CONNECTION, POWERED ROBOT TESTING, MOTION, WALKING OR ENERGIZATION |
contact map · cut list · traveler · STEP · primary sources